We understand that many organizations may want to centrally manage their network resources to simplify security, updates, and authentication.
While we cannot speak on all the potential controls or customizations your organization may push through group policy, it is perfectly fine to join your specular microscope's computer to a network domain.
However, we recommend that your users still log into the local "user" account when using CellChek devices, as they contains numerous customizations for a smooth user experience. Otherwise, domain users will need desktop shortcuts setup to run the software easily.
For exporting records to a network location, for EMR purposes, we simply recommend using a mapped drive and saving the credentials. Adding the device to a network domain is not necessary for this.
We also recommend whitelisting a number of files and folders to prevent any performance issues or false positive detections that may remove critical files required to run your Konan product's software.
Below is a list of file, folder, and hardware device exclusions that should be provided to your IT security personnel to be added to the solution's Whitelist / Allowlist.
Please also note, the folders that require special permissions.
CellChek20 and 20+ models:
|
Folders: C:\Bin C:\NSP17 D:\AlignmentLog D:\Data D:\EMR D:\Filing D:\Gateway D:\Log |
Processes: CellChek_X8.exe DSCameraControl.exe Eshell.exe Initialize221.exe startCCX8.exe
|
Hardware Devices: Basler USB3 Vison Cameras -Basler dart USB3 Vision Camera -Basler USB3 Vison Camera (depending on CC20 version) Ports (COM & LPT) -USB Serial Port (COM#) Universal Serial Bus Controllers -uEye UI-327x Series (depending on CC20 version)
|
Please note: all Windows user accounts that will be using the CellChek software, must also have full read/write/delete rights to the folders listed above (domain users).
CellChek XL / SL models:
|
Folders:
C:\CellChek C:\Konan_Drivers (or C:\Konan Drivers) C:\Gateway for Konan System (for DICOM software only) C:\Watch (for DICOM software only)
|
Processes:
CellChek.exe Konan Analysis Suite.exe
|
Hardware Devices: Hauppauge USB Live 2 Ports (COM & LPT) -USB Serial Port (COM#) Universal Serial Bus Controllers |
Please note: all Windows user accounts that will be using the CellChek software, must also have full read/write/delete rights to the folders listed above (domain users).
Konan Medical USA
Customer Service and Technical Support:
Available 6:30 AM–5:00 PM Mon-Fri Pacific Time